Volumetric modulated arc therapy treatment planning assessment for low-risk prostate cancer in radiotherapy
Hafiz Zulkafal, Muhammad Khan, Muhammad Ahmad, Muhammad Akram, Saeed Buzdar, Khalid Iqbal

Aims: This research study was carried out for dose verification of volumetric modulated arc therapy (VMAT) plan in the patients of low-risk prostate cancer using different indices in radiotherapy. Materials and Methods: Twenty patients with low-risk prostate cancer after histopathological confirmation were included which were divided into two groups with prescribed dose of 7400 and 6000 cGy according to their disease and 3 mm sliced for each patient by computed tomography was obtained for planning. To evaluate the dose conformity and its distribution using VMAT plans such as homogeneity index (HI), radial dose HI, moderate dose HI, coverage, and uniformity index for each patient. These plans were constructed for each patient separately using Eclipse Treatment planning software. Statistical Analysis Used: Statistical Package for the Social Sciences statistical software is used to calculate mean standard deviation and standard error of all these indices. Results: Obtained results and findings enabled this research work to conclude that VMAT is a better treatment technique for the patients of low-risk prostate cancer with uniform dose distribution. Conclusion: It has been suggested that VMAT with Eclipse Treatment planning system is a good treatment modality to treat the patients of low-risk prostate cancer with better results.

Dose distribution, eclipse treatment planning software, indices in radiotherapy, prostate cancer, volumetric modulated arc therapy plan verification

Introduction

Radiotherapy is one of the most adaptive methods of cancer treatment in the world. Ionizing radiations are being used to destroy cancerous cells in human body. Basically, in radiotherapy, only tumors cells are required to receive the prescribed dose while normal tissues and nearby organs at risk be secured. This requires an excellent care in planning for dose distribution and delivery. On the basis of anatomical structure, accurate treatment plan is made taking geometric and dosimetric consideration. For tumor treatment geometric accuracy of 2–3 mm in dose delivery is essentially needed.[1] In conformal dose distribution, 5%–7% dosimetric differences in target volume can harm healthy cells and considerable change of tumor control.[2],[3] Treatment plans verification and exact delivery of dose in quality assurance programs prevent errors and accidents and provide high-level accuracy in radiotherapy. Advancement in radiotherapy, especially in external beam radiotherapy (EBRT) and modern treatment planning systems have revolutionized the treatment of tumor at any site. In EBRT, volumetric modulated arc therapy (VMAT) is the best example. VMAT technique is the advanced form of intensity modulated radiotherapy (IMRT). VMAT deliver dose with control dose rate and multi-leaf collimator with the adjustment of gentry's rotation speed.[4] When gentry rotate about the isocenter the beam with multiple or single arc (SA) is delivered in this technique. The aim of this article is to study the different parameters for dose verification of VMAT double arc (DA) treatment plans for prostate cancer. Prostate cancer is mostly diagnosed in gents and is second commonly occurring cancer after skin cancer in the developed countries.[5],[6] VMAT in EBRT provides the best opportunity due to conformal dose distribution to tumor and decrease the dose to other parts of body.[7],[8],[9] VMAT is preferred in prostate cancer treatment over IMRT as it consumes lower number of monitor units and same dose distribution like IMRT.[7],[8],[9] Many studies have been carried out to understand the clinical application of RapidArc for prostate cancer treatment. In all these studies variations in the outputs were observed because of planning techniques, shape of the target, objective defined, and target volume margins.[10],[11],[12],[13],[14],[15],[16],[17],[18] Two types of plans are utilized for prostate cancer patients having SA and DA. Sze et al.[11] gave the comparison of 14 prostate cancer patients between SA and DA and proved that DA technique is better than SA due to better homogeneity, conformity, and lowest hotspot. This study also explains rectal dose using SA was maximum than DA techniques. Pengpeng Zhang with his colleagues also compared the VMAT plan and Standard IMRT plan and showed that VMAT treatment decreases the Beam-on time up to 55% while keeping the dosimetric quality same.[13] James L Bedford and his Companion also investigated VMAT plans for different site included prostate site and proved that VMAT has high-quality dose delivery in shorter time as compared to IMRT.[19] Currently, a prostate cancer RapidArc plan with two full arcs has been considered to be the standard plan for the improvement in prostate cancer treatment.

Materials and Methods

In this research project, twenty patients with low-risk prostate cancer were included after confirmation by histopathological test. Computed tomography of 3 mm sliced was obtained of all patients for consideration in simulation and planning to investigate the dose conformity and distribution by different parameters for VMAT plans. All these patients were treated by 6MV photon beam of Clinic DHX linear accelerator (Varian Medical Systems Inc., Palo Alto, CA). VMAT treatment Plans were made on the basis of computed tomographs relevant to each patient separately using Eclipse Treatment Planning Software (ARIA 11) (Varian Medical Systems Inc., Palo Alto, CA). This study involves only low-risk prostate cancer patients with defined clinical target volume (CTV) [Figure 1].{Figure 1}

In all directions, the addition of 1 cm to CTV planning target volume (PTV) was considered excluding posterior. 7 mm margin from CTV is taken for PTV to minimize dose of rectum. According to radiation therapy oncology group (RTOG-0815) protocol [20] plans are drawn for Critical organs such as femoral heads, rectum, and bladder. A cumulative dose-volume histogram (DVH) was obtained from these plans and used to calculate respective dose values and volumes for each Patient separately for 6-MV beam. The homogeneity index (HI) and dose coverage defined by RTOG in 1993 was used for analyzing the reliability of plan.[21] Radical HI and moderate dose homogeneity indices defined by Oliver in 2007[22] are utilized in this research study for VMAT plan verification. Uniformity index (UI) used by Chitapanarux I in 2015[23] is also calculated, and all these indices are calculated according to defined relations.

Following relations have been used for calculation of these indices are described by their users.

HI = MD/PD (MD = maximum dose used in DVHs and PD = prescribed dose).

Coverage = Dmin/PD (Dmin= minimum dose described in the DVHs).

Radical dose HI (rDHI) = Dmin/Dmax(Dmax= MD).

Moderate dose HI (mDHI) = D95/D5(D95= dose at 95% and D5= dose at 5% of target volume)

UI = D5/D95.

The cumulative DVHs are generated automatically by the treatment planning software from where all these values are calculated.

Results

Two values of the prescribed dose are used in this project for PTV that is 7400 cGy and 6000 cGy. A total of 20 cases were analyzed, 9 were treated by 7400 cGy PTV and 11 patients were planned for 6000 cGy for PTV. Only the dose values of PTV are considered to calculate these indices which are shown in the graph by red line and the other lines showing the dose delivery to different nearby organs.

D5 and D95 are the doses at 5% and 95% volumes calculated from DVHs carefully. Dmax and Dmin are the maximum and minimum doses used and defined in DVHs that can also be seen in [Figure 2]. These values are defined separately for each case, which are different from each other listed in [Table 1].{Figure 2}{Table 1}

D5 ranges from minimum value of 6190 to maximum value of 7950 while the range of D95 is from 5780 to 7470. Dmax contains the values between 6412.7 and 8083.4 and Dmin having the values between 5289.7 and 6832.0.

[Table 2] contains calculated values of indices (HI, coverage, UI, mDHI, and rDHI) for each case. Standard deviation and mean standard error is calculated for twenty cases, first nine cases having the PD 7400 cGy and last eleven cases having the PD 6000 cGy separately.{Table 2}

Discussion

In this research study, findings of 20 patients were analyzed, 9 of them were treated by PD of 7400 cGy and 11 were treated by 6000 cGy. In both cases, HI occupies the range of 0.07 from 1.04 to 1.11. Lower values of homogeneity reflect better dose distribution.[24] Standard deviation of homogeneity comes to be 0.02 and mean standard error is 0.004. These are very low and nonconsiderable values. Dose coverage for all cases has values 0.88–0.92 with range of 0.4. With standard deviation of 0.02 and mean standard error of 0.002. Such lower values are in satisfactory limit with UI range of 0.05 from 1.04 to 1.09 with standard deviation of 0.02 and mean standard error of 0.004. Moderate HI for all cases occupied the range of 0.06 from 0.90 to 0.96 with standard deviation of 0.02 and mean standard error of 0.003. Radical HI has the range of 0.06 from 0.81 to 0.87 with standard deviation of 0.02 and mean standard error of 0.004. Same range of moderate dose HI and radical dose HI reflects the uniform distribution of dose.[25]

In both cases, it is observed for UI, medical dose HI, and dose Coverage that with increasing PD the standard deviation and mean standard error decreases however for HI and radical dose HI it is observed that with increasing PD the standard deviation and mean standard error also increase. Regardless of this trend, it is seen that all the values are so smaller fractions that these are insignificant. All these indices show the values in the satisfactory range defined by RTOG and no single entry shows any deviation from the defined limit. In addition, puniness in the values of standard deviation of these indices in signifies the changes in the delivery of dose.

Conclusion

This research work aimed for dose verification in low-risk prostate cancer cases using different indices of VMAT plan and on the basis of obtained data it has been concluded that VMAT is a better treatment technique for low-risk prostate cancer cases with uniform dose distribution. Furthermore, it is also suggested that the combination of LINAC DHX linear accelerator and Eclipse Treatment planning software is a good treatment modality to treat the patients of low-risk prostate cancer with better efficacy.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

References

Källman P, Agren A, Brahme A. Tumour and normal tissue responses to fractionated non-uniform dose delivery. Int J Radiat Biol 1992;62:249-62.

Ringborg U, Bergqvist D, Brorsson B, Cavallin-Stahl E, Ceberg J, Einhorn N, et al. The Swedish council on technology assessment in health care (SBU) systematic overview of radiotherapy for cancer including a prospective survey of radiotherapy practice in Sweden 2001 – Summary and conclusions. Acta Oncol 2003;42:357-65.

Verellen D, De Ridder M, Linthout N, Tournel K, Soete G, Storme G, et al. Innovations in image-guided radiotherapy. Nat Rev Cancer 2007;7:949-60.

Rozario, Timothy Michael. Markerless lung tumor tracking algorithms for image guided radiation therapy. Lahore, Pakistan: Shaukat Khanum Memorial Cancer Hospital and Research Center; 2015.

Torre LA, Bray F, Siegel RL, Ferlay J, Lortet-Tieulent J, Jemal A, et al. Global cancer statistics, 2012. CA Cancer J Clin 2015;65:87-108.

Thun MJ, DeLancey JO, Center MM, Jemal A, Ward EM. The global burden of cancer: Priorities for prevention. Carcinogenesis 2010;31:100-10.

Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008;35:310-7.

Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Losasso T, et al. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 2008;72:575-81.

Tang G, Earl MA, Luan S, Wang C, Mohiuddin MM, Yu CX, et al. Comparing radiation treatments using intensity-modulated beams, multiple arcs, and single arcs. Int J Radiat Oncol Biol Phys 2010;76:1554-62.

Palma D, Vollans E, James K, Nakano S, Moiseenko V, Shaffer R, et al. Volumetric modulated arc therapy for delivery of prostate radiotherapy: Comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys 2008;72:996-1001.

Sze HC, Lee MC, Hung WM, Yau TK, Lee AW. RapidArc radiotherapy planning for prostate cancer: Single-arc and double-arc techniques vs. Intensity-modulated radiotherapy. Med Dosim 2012;37:87-91.

Yoo S, Wu QJ, Lee WR, Yin FF. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vesicles and lymph nodes. Int J Radiat Oncol Biol Phys 2010;76:935-42.

Zhang P, Happersett L, Hunt M, Jackson A, Zelefsky M, Mageras G, et al. Volumetric modulated arc therapy: Planning and evaluation for prostate cancer cases. Int J Radiat Oncol Biol Phys 2010;76:1456-62.

Betzel GT, Yi BY, Niu Y, Yu CX. Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT? Med Phys 2012;39:5882-90.

Kjaer-Kristoffersen F, Ohlhues L, Medin J, Korreman S. RapidArc volumetric modulated therapy planning for prostate cancer patients. Acta Oncol 2009;48:227-32.

Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M, et al. Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother Oncol 2009;93:259-65.

Wolff D, Stieler F, Welzel G, Lorenz F, Abo-Madyan Y, Mai S, et al. Volumetric modulated arc therapy (VMAT) vs. Serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother Oncol 2009;93:226-33.

Shaffer R, Morris WJ, Moiseenko V, Welsh M, Crumley C, Nakano S, et al. Volumetric modulated Arc therapy and conventional intensity-modulated radiotherapy for simultaneous maximal intraprostatic boost: A planning comparison study. Clin Oncol (R Coll Radiol) 2009;21:401-7.

Bedford JL. Treatment planning for volumetric modulated arc therapy. Med Phys 2009;36:5128-38.

Martinez AA. RTOG 0815 Protocol Information. Radiation Therapy Oncology Group [Electronic]; 2012. Available from: http://www.rtog.org/ClinicalTrials/ProtocolTable/StudyDetails.aspx?study=0815. [Last accessed on 2014 Jun 23].

Shaw E, Kline R, Gillin M, Souhami L, Hirschfeld A, Dinapoli R, et al. Radiation Therapy Oncology Group: Radiosurgery quality assurance guidelines. Int J Radiat Oncol Biol Phys 1993;27:1231-9.

Oliver M, Chen J, Wong E, Van Dyk J, Perera F. A treatment planning study comparing whole breast radiation therapy against conformal, IMRT and tomotherapy for accelerated partial breast irradiation. Radiother Oncol 2007;82:317-23.

Chitapanarux I, Tharavichitkul E, Nobnop W, Wanwilairat S, Vongtama R, Traisathit P, et al. Acomparative planning study of step-and-shoot IMRT versus helical tomotherapy for whole-pelvis irradiation in cervical cancer. J Radiat Res 2015;56:539-45.

Iqbal K, Isa M, Buzdar SA, Gifford KA, Afzal M. Treatment planning evaluation of sliding window and multiple static segments technique in intensity modulated radiotherapy. Rep Pract Oncol Radiother 2012;18:101-6.

Ibbott GS, Hanson WF, O'Meara E, Kuske RR, Arthur D, Rabinovitch R, et al. Dose specification and quality assurance of Radiation Therapy Oncology Group protocol 95-17; A cooperative group study of iridium-192 breast implants as sole therapy. Int J Radiat Oncol Biol Phys 2007;69:1572-8.

© Clinical Cancer Investigation Journal
Online since 01 December, 2011
Creative Commons License 
ISSN Print: 2278-1668, Online: 2278-0513